PLG-2000SUP Abnormal waveform ECG signal simulator PVC,VPC, ECG
signal generator, Arrhythmia ECG simulator
Introduction:
This model simulator suitable for ECG machine manufacturer, or ECG
kind product manufacturer. It can be used to adjust machine, and
use for ECG machine development.
Feature:
Rechargeable Lithium battery
10 pcs multi-functional ECG connector, can be connect with
different type ECG machine cable or patient monitor cable
Battery power indicate light
Low battery power automatic turn off function, protect the Lithium
battery
Description
The waveform type description ( using the Mindray PM8000 monitor
detection )
1, waveform code 1: normal ECG waveform
The data code at 60 is mean :heart rate 60 bpm
The frequency range is 10-200 BPM and wave amplitude is fixed, it
will not be affected by the amplitude setting.
2, waveform code 2: It is represent triangular waveform waveform
code
Data code at 75 : it is mean heart rate 75bpm
Frequency range: 10-400bpm
Amplitude range: 0.1-4mV, the amplitude can be regulated in
regulate setting place
3, waveform code 3: Square waveform
Data code at 10 : it is mean 1Hz
Frequency range: 0.1-10Hz
Amplitude range: 0.1-4mV, the amplitude can be regulated in
regulate setting place
4, Waveform code 4 :Sinusoidal waveform
Data code at 10 : it is mean 10Hz
Frequency range : 1-100Hz
Amplitude range: 0.1-4mV, the amplitude can be regulated in
regulate setting place
5, Waveform code 5: QRS wave waveform
You can change the amplitude of T wave, QRS wave amplitude, width
of QRS wave, QRS wave frequency
To change these parameters through the combination of keys
6.Waveform code at 6: Respiratory waveform
Data code at 30: it is mean 30bpm
Respiratory waveform frequency adjustment range 10-120 times
7, Waveform code 7: pacing signal ,the pacing pulse signal of
positive and negative direction
Data code at 20 ( It is represent 2mS )
The frequency is 60bpm
Amplitude range: 0.1-4mV, the amplitude can be regulated in
regulate setting place
Through the combination of keys to switch the positive and negative
direction
You can change the pulse width, the range is 0.5-2mS
8, Waveform code 8: QRS wave of front pacing signals
Front pacing signal is positive and negative pulse waveform,
through the combination of keys to switch
The width of pacing signal pulse can be adjusted, the range is
0.5-2mS
Amplitude range: 0.1-4mV, the amplitude can be regulated in
regulate setting place
9, Waveform code 9: adjust the waveform amplitude
Data code at100 : it is represent 1Mv
400: it is represent 4Mv
10, waveform code at A:
Data code | bpm | Symptom |
Data code at 1 | 75bpm | Normal |
Data code at 2 | 140bpm | Tachycardia |
Data code at 3 | 35bpm | Bradycardia |
Data code at 4 | 200bpm | T supraventricular tachycardia |
Data code at 5 | 80(73)bpm | Missed beat |
Data code at 6 | 14 bpm | Agony rhythm |
Data code at 7 | 0 bpm | Cardioplegia |
11, Waveform code at B:
Data code | bpm | Symptom |
Data code 1 | 75 bpm | Normal |
Data code 2 | 82 bpm | NSR with PAC |
Data code at 3 | 82 bpm | Coarse atrial fibrillation, arhythmia |
Data code at 4 | 80 bpm | premature ventricular contraction |
Data code at 5 | 80 bpm | Continuation 2 premature ventricular contraction |
Data code at 6 | 80 bpm | Continuation 3 premature ventricular contraction |
Data code at 7 | 80 bpm | Continuation 4 premature ventricular contraction |
Data code 8 | 80 bpm | Premature beat |
Data code 9 | 80 bpm | 2 continuation premature beat |
Data code 10 | 80 bpm | 4 continuation premature beat |
Data code 11 | 80 bpm | 1 premature beat,1 normal waveform |
Data code 12 | 80 bpm | All premature beat(VF/VT) |
12, waveform code at C:
Data code | bpm | Symptom |
Data code at 1 | 60 bpm | Degree block |
Data code at 2 | 35 bpm | Degree block 30 bpm |
Data code at 3 | 31 bpm | Degree block 31 bpm |
Data code at 4 | 60 bpm | L bnd1 branch block |
Data code at 5 | 60 bpm | R bnd1 branch block |
13, waveform code at D:
Data code number | bpm | Symptom |
Data code at 1 | (80-83) bpm | ATRIAL PAC |
Data code at 2 | 60 bpm | Pvc1 early Lv focus |
Data code at 3 | 60 bpm | Pvc1 early Rv focus |
Data code at 4 | 60 bpm | Nodal pnc |
Data code at 5 | 60 bpm | PVCs bigeminy |
Data code at 6 | 60 bpm | PVCs trigeminy |
Data code at 7 | 80 bpm | R ON T |
Precise life group limited
Precise Life——Global Medical Equipment Accessories Solution Expert
Since its establishment in 2010, Precise Life has been committed to providing **high-quality medical equipment accessories and parts** to global medical institutions, distributors and service providers, helping global medical equipment to operate efficiently and protecting the lives and health of patients. With our professional technical capabilities, strict quality standards and efficient service network, we have become a trusted partner in the medical industry in more than 60 countries.
Our products cover **all categories of medical equipment
accessories and parts**, focusing on providing key technical
support for the following equipment:
- **Ventilator accessories**: flow sensor, pressure sensor,
motherboard, filter module, breathing circuit components, etc.
- **Anesthesia machine accessories**: various boards, evaporator
modules, gas valves, pressure sensors, breathing airbags, etc.
- **Monitor accessories**: ECG lead wires, blood oxygen probeSpo2
sensos, blood pressure cuffs, display modules, etc.
- **Medical ultrasound instrument accessories**: ultrasound probes,
connecting cables, touch screens, etc.
At the same time, we provide **sensors, circuit boards, connecting
cables, mechanical components** and other diversified precision
parts, suitable for mainstream brand equipment models such as GE,
Philips, Siemens, etc.
All products have passed ISO 13485 medical device quality management system certification, and comply with international standards such as FDA and CE to ensure safety and reliability.
Relying on the global logistics supply chain, we provide efficient logistics and flexible order services to meet urgent needs.
In addition to the supply of accessories, we provide **one-stop equipment maintenance services**, with a team of senior engineers to diagnose, replace parts and optimize performance, repair parts, and reduce operation and maintenance costs.
Precise Life has always been committed to "**keep medical equipment running and reduce health maintenance costs**" and promote the efficient use of global medical resources through innovative technologies and services. In the future, we will continue to deepen our roots in the field of medical accessories and contribute to protecting human health.
**Contact us**
Whether you need standard accessories procurement or maintenance
support, the Precise Life team is always ready to provide you with
professional answers!