SpO2 Medical simulator with ECG simulator function SpO2 simulator with 10 lead
SpO2 Function introduction:
1. External imitate finger,you can connect to any SpO2 test instruments easily; |
2. It is a multi-functional optical simulator, built-in frequently-used BCI,Mindray,Massi-mo, wave-forms curve; |
3. The Oximetry simulate range:80%,85%,90%,98%, these 4 points value detection calibration,tolerance<=1%; 60%,65%,70%,75%,these 4 points value detection calibration, tolerance<2%; |
4. Pulse rate simulate range:30,60,80,100,120,160,180,240 bpm total 8 test points, tolerance<=1bpm; |
ECG function introduction:
1. Normal ECG waveform, 12-lead different ranges and types synchronous waveform; |
2. Plus and reserve direction heart rate detection waveform; |
|
4. Sinusoidal wave, measure amplitude-frequency characteristic |
|
6. The Respiration rate of breathing waveform can be changed+ ECG waveform; |
7. You can set the ECG signal amplitude; |
Waveform code introduction:
1. Oxygen saturation values, set the SpO2 rate of simulator;
2. Pulse rate value, set the simulator pulse rate value;
3. Select the data curve of Oxygen saturation;
4. Normal ECG waveform, 12-lead different ranges and types
synchronous waveform;
5. Plus-minus triangular waveform;
6. Square wave, used the square wave to measure scan speed;
7. Sinusoidal wave, measure amplitude-frequency characteristic
8. Can change the T-wave amplitude, heart rate value, R-wave
amplitude, and R-wave width of analog QRS waveform;
9. Simulate Respiration waveform;
A. Set the amplitude of signal;
Note:
1.The simulator will automatically generate a value 98 of blood
Oxygen, 80 of pulse rate, blood Oxygen parameter data of curve,
and code as 4 normal ECG waveform, and respiration rate as 15bpm
respiratory waveform.
2. The imitate finger part, against the LED of the Spo2 sensor; and
the imitate finger part, against the SpO2 sensor PD; and make sure
that the white color window of imitate finger stay in line with the
PD of SpO2 sensor, or it will lead to invalid data.
3. The simulator will automatically generate lead II amplitude at
1mV,and in the heart rate detection wave, square wave, sinusoidal
wave,
QRS wave,all amplitude of lead II is 1mV.
4. When the battery falls lower than a certain value, the contents
of the digital display will flash to indicate low battery, and then
continue
to work only about four hours. Please note that at this moment the
ECG waveform amplitude will decrease.
5. The ECG part connection to ECG machine: RA-R(right arm),
LA-L(left arm), LL-F(left leg),RL-RF(right leg), C1-C6(chest lead).
6. The ECG part connection to patient monitor:RA-right arm(white),
LA-left arm(black), LL-left leg(red), RL-right leg(green), C1-C6
chest lead(brown).
7. ECG part 3 leads connection: RA-right arm(white), LA-left
arm(black), LL-left leg(red).
8. Different identify corresponding connection:L-LA, R-RA,
RF(N)-RL, F-LL, C-V.
Key description:
There are 4 keys, followed by selection keys, increase key,
reducing key, the Enter key, in addition to a set of key
combinations
Select Key:
This key is used to select the parameter you want to change; there
are 4 LED tubes to display the four codes represent what is
displayed,
The 1st LED representative waveform code, LED 2-4 representatives
the parameter (2 is the value of one hundred, 3 representatives of
ten, 4 representatives of bits)
Increase key:
Use it to increase the parameter
Reducing key:
Use it to reduce the parameter
Enter key:
Use it to confirm the parameter you selected
Combination key:
Press the select key and Enter key at the same time, release the
Enter key, the release the select key,
double-click the Enter key again, the combination key was selected
successfully, and it will display different content.
The operation ways of each code:
1. Blood Oxygen value:
1) The blood Oxygen value setting at:80%,85%,90%,98%, these 4
points value detection calibration,tolerance<=1%;
At 60%,65%,70%,75%,these 4 points value detection calibration,
tolerance<2%;
2) Initial value:98%; It can be select and set the blood oxygen
value with increase key and reduce key directly.
2. Pulse rate value:
1) Pulse rate value setting at: 30,60,80,100,120,160,180,240 bpm
total 8 test points;
2) Tolerance<=1 bpm;
3) Initial value is 80 bpm,It can be select and set the pulse rate
value with increase key and reduce key directly.
3. Waveform data curve selection:
Data 1 is BCI curve, data 2 is curve.
4. Normal ECG waveform:
* Heart rate setting range: 10-200bpm (Initial value: 60 bpm)
* Signal amplitude is fixed
5. Plus-minus triangular waveform
* Frequency range: 10-300bpm (Initial value: 75 bpm)
* Amplitude range: 0.1-4Mv (10:0.1mV, 400:4mV)
* 2 modes, Mode 1: Positive waveform; mode II: negative waveform;
Use combinations key to choose.
6. Square wave
* Frequency range: 0.1Hz-10Hz (10:0.1Hz, 100:10Hz); (Initial value:
1 Hz )
* Amplitude range: 0.1-4Mv (10:0.1mV, 400: 4mV);
7. Sinusoidal wave
* Frequency range: 1-100Hz (Initial value: 25 Hz)
* Amplitude range: 0.1-4Mv (10:0.1mV, 400:4mV)
8. Analog QRS-T waveform:
This waveform have 4 operating modes,
Model 1: the T-wave amplitude can be changed;
Model 2: Set the waveform frequency;
Mode 3: setting the amplitude of QRS waveform;
Model 4: Set the width of QRS wave;
the operation mode can change with combination keys.
* T wave amplitude setting range: 0.1mV-2mV; (Initial value: 10)
* Waveform frequency setting range: 20-300bpm; (Initial value: 75)
* QRS waveform amplitude range: 0.1mV-2mV;(Initial value: 1mV)
* QRS waveform width range: 10ms-150ms;(Initial value: 80ms)
9. Simulate respiration wave:
* Frequency range: 10-100time/min; (Initial value: 15bpm)
* Note: the respiration lead is RA-LL, baseline impedance is 1k.If
the patient monitor respiration lead is other lead, please set the
parameter of monitor or change the lead connect way.
A. Signal amplitude setting:
* Amplitude range: 0.1mV-4mV (10:0.1mV, 400:4mV); (Initial value:
100)
* Note: this amplitude setting will affect the 5,6,7 waveform
amplitude.
Precise life group limited
Precise Life——Global Medical Equipment Accessories Solution Expert
Since its establishment in 2010, Precise Life has been committed to providing **high-quality medical equipment accessories and parts** to global medical institutions, distributors and service providers, helping global medical equipment to operate efficiently and protecting the lives and health of patients. With our professional technical capabilities, strict quality standards and efficient service network, we have become a trusted partner in the medical industry in more than 60 countries.
Our products cover **all categories of medical equipment
accessories and parts**, focusing on providing key technical
support for the following equipment:
- **Ventilator accessories**: flow sensor, pressure sensor,
motherboard, filter module, breathing circuit components, etc.
- **Anesthesia machine accessories**: various boards, evaporator
modules, gas valves, pressure sensors, breathing airbags, etc.
- **Monitor accessories**: ECG lead wires, blood oxygen probeSpo2
sensos, blood pressure cuffs, display modules, etc.
- **Medical ultrasound instrument accessories**: ultrasound probes,
connecting cables, touch screens, etc.
At the same time, we provide **sensors, circuit boards, connecting
cables, mechanical components** and other diversified precision
parts, suitable for mainstream brand equipment models such as GE,
Philips, Siemens, etc.
All products have passed ISO 13485 medical device quality management system certification, and comply with international standards such as FDA and CE to ensure safety and reliability.
Relying on the global logistics supply chain, we provide efficient logistics and flexible order services to meet urgent needs.
In addition to the supply of accessories, we provide **one-stop equipment maintenance services**, with a team of senior engineers to diagnose, replace parts and optimize performance, repair parts, and reduce operation and maintenance costs.
Precise Life has always been committed to "**keep medical equipment running and reduce health maintenance costs**" and promote the efficient use of global medical resources through innovative technologies and services. In the future, we will continue to deepen our roots in the field of medical accessories and contribute to protecting human health.
**Contact us**
Whether you need standard accessories procurement or maintenance
support, the Precise Life team is always ready to provide you with
professional answers!